Feature |
Description |
| Measuring principle |
Dual beam sphere - automatic specular port |
| Measuring geometry |
Diffuse illumination 8° viewing in conformance with CIE publication No. 15.2 Colorimetry. |
| Light source |
Pulsed xenon, optional UV calibrator |
| Spectral range |
360 to 700nm (depending on model) |
| Effective bandwidth |
10nm |
| Wavelength resolution |
2 nm |
| Spectrometer principle |
Concave holographic grating |
| Detector |
Proprietary active pixel dual 256 element diode array |
| Sphere diameter |
51mm/2 in |
| Measuring range |
0 to 200% reflectance |
| Measurement time |
< 2.5 seconds |
| Aperture sizes |
LAV 15 mm 11 mm |
|
SAV 10 mm 6.5 mm |
|
USAV 6.5 mm 2.5 mm |
| Repeatability |
0.03 CIELAB E maximum white tile, 2 flash measurement |
| Reproducibility |
0.15 average CIELAB E, 12 BCRA tiles, 0.25 CIELAB E maximum |
| Battery |
Replaceable lithium ion |
| Battery life |
>2,000 measurements |
| Operating environment |
5° to 40° C up to 85% RH, non-condensing |
| Interface |
RS232 and USB |
Feature |
Description |
| Size |
9.7 cm x 9.5 cm x 24.6 cm / 3.85 in x 3.75 in x 9.7 in |
| Display |
High resolution, 320x320 color screen |
|
5.4 cm x 5.4 cm |
|
2.125 in x 2.125 in |
| Weight |
1 Kg/2.2 lbs |
| PDA |
PalmOne™ Tungsten™ with Palm® OS (included) |
| Accessories |
Soft carry bag containing: Power supply and country cord, Stapler foot for sample alignment, RS232 and USB cables, White tile, Black trap |